Features
- The smallest compact vital signs monitor footprint in the world.
- NIBP adopts two patents in NIBP, SAWAA (Self-Adaptive-Waveform-Amplitude-adjustment) and SAFD (Self-Adaptive-fast-Deflation) algorithms similar to Petmap method.
- M-Series monitors configuration
- M6 = ECG+ Sp02 + NIBP + TEMP
- M6E = Sp02 + NIBP +TEMP+ MainStream EtC02
- M6SE = Sp02 + NIBP +TEMP+ SideStream EIC02
- Two measurement modes: Spot or Monitoring modes
- Data storage capabilities – historical data view, trendgraph view.
- Monitoring mode – continuous recording, 48 hours of measurement data.
- Spot mode – 100 user ID’s, each user can store 200 data sets.
- Audible and visual alarms, with upper and lower alarm limit adjustment.
- Lithium Battery.
Performance Specifications
NIBP
- Wid NIBP measurement range
Large Animal | Small Animal | |
SYS | 40 – 270mmHg | 40 – 220mmHg |
MEAN | 20 – 230mmHg | 20 – 165mmHg |
DIA | 10 – 210mmHg | 10 – 150mmHg |
- NIBP measurement accuracy: ±3mmHg
- Pulse rate measurement accuracy: ≤ 2% (≥ 100 BPM) or 2 beats/min (<100 BPM)
Oxygen Saturation (%SpO2)
- Measurement range: SpO2: 0-100%
- PR: 0-350bpm
- Perfusion index: 0.05% – 20%
Temperature
- Measurement range: 25 – 45°C
- Accuracy: ±0.1%
End Fidal CO2 (EtCO2)
- Measurement range: 0 – 150mmHg (0-19.7%) (0-20kPa) RR: 3-150BPM
- Accuracy range:
ETCO2 Range | Accuracy |
0 – 40mmHg | ±2mmHg |
41 – 70mmHg | ±5% of Reading |
71 – 100mmHg | ±8% of Reading |
101 – 150mmHg | ±10% of Reading |
- RR Accuracy: ±1BPM
ECG
- Lead type:
- 3-Leads or 5-Leads.
- Waveform selection:
- 3-Leads: i, ii, iii
- 5-Leads: i, ii, iii, AVR, AVR, AVF, V
Technical Specifications
Display |
4.3” Color TFT |
Resolution |
320*480 |
Indicator |
Alarm indicator, Power Indicator |
Trace |
1 plethysmogram wavefrom |
Alarm |
Sensor off, Low power, Checking Adapter, ECG cable off Etc. |
Modes |
Visual and Audio |
Application |
Large and small animals | ||
Size |
180mm*100mm*80mm |
Net Weight |
250g |